BS EN 13060:2025
BSI · 2025-08-31

Sterilizers for medical purposes. Small steam sterilizers. Requirements and testing

试验 ICS 11.080.10

1Key Takeaways

1 Scope This document specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids. This document applies to automatically controlled sma…

2Expert Interpretation

BS EN 13060:2025 specifies the design, performance requirements and test methods for small steam sterilizers for medical use, covering sterilization cycles of types B, N and S, ensuring steam penetration and sterilization reliability, and is suitable for dental, cosmetic and other places.

Standard Background and Technological Evolution

BS EN 13060:2025 "Medical sterilizers — Small steam sterilizers — Requirements and tests" is the latest standard published by the European Committee for Standardization (CEN), replacing the 2014 version. This standard aims to regulate small steam sterilizers with a volume of less than 60 liters that cannot accommodate a standard sterilization module (300 mm × 300 mm × 600 mm), widely used in general practice, dentistry, podiatry, beauty, and tattooing. Compared to the previous version, the 2025 version adopts the framework of ISO/TS 22421:2021 in structure, updates normative references, and adjusts terminology to align with EN ISO 11139 and EN ISO 17665. Technological evolution is mainly reflected in a clearer distinction between process variables (temperature, time) and cyclic variables (pressure), the introduction of independent monitoring system requirements, and more stringent drying performance indicators.


Scope and Core Definitions

This standard applies to automatically controlled small steam sterilizers, whose steam can be generated by an electric heater or supplied by an external system. The sterilizer must meet the following conditions: **chamber volume less than 60 liters**, and cannot accommodate standard sterilization modules. The primary sterilization targets are medical devices and materials that come into contact with blood or bodily fluids. The standard does not cover the sterilization of liquids or pharmaceuticals, nor does it address processes involving formaldehyde or ethylene oxide. Core terms include: **sterilization cycle**, **process challenge device (PCD)**, **biological indicator**, **chemical indicator**, etc. Among these, **small steam sterilizer** specifically refers to equipment with a volume less than 60 liters that cannot accommodate 300mm × 300mm × 600mm modules.

Comparison of Sterilization Cycle Types

< Type N Type S
Cycle TypeDescriptionApplicable LoadsTypical Tests
Type BSterilizes all standard test loads, including solid, porous, and tubular itemsWrapped or unwrapped solid, porous, and tubular itemsAir leakage, dynamic pressure, cavity, solid load, narrow cavity, porous load, etc.
Applicable only to unwrapped solid items Unwrapped solid instruments (ready to use) Dynamic pressure, cavity, solid load (unwrapped)
Specified by the manufacturer, including at least unwrapped solid and one other type Specific load (e.g., porous, lumen, single-layer wrapping, etc.) Select the appropriate test according to the claimed load type
Each cycle type demonstrates its performance through different combinations of tests. For example, Type B cycles must pass the steam penetration test of narrow-cavity tubing (using PCD conforming to EN ISO 11140-6), while Type N cycles only need to pass the solid load test.


Key Design Requirements

The standard sets forth detailed requirements for the design of sterilizers. The chamber must have an air leak test capability, with a leakage rate not exceeding 0.13 kPa/min (when using the vacuum stage). Pressure vessels must comply with the EN 13445 series standards. A door interlock system ensures that the door cannot be opened during sterilization and can be safely restored in case of failure. The control and recording system must be independently monitored: at least two independent temperature measurement chains (or one temperature and one pressure measurement chain) are required, and anomalies are detected by comparing theoretical steam temperature with actual temperature. Software must be validated according to EN 62304, and cybersecurity must be considered. Regarding service facilities: Electrical supply must meet ±10% voltage fluctuations; water supply quality must meet the limits in Annex F (e.g., conductivity ≤15 μS/cm); and the non-condensable gas content of external steam must not exceed 3.5%. Performance Testing Methods and Indicators: The standard specifies various types of tests: type tests, factory tests, and installation tests. Core performance indicators include:
- **Sterilization Temperature Range:** The temperature must not be lower than the set value, and must not exceed the set value by more than 3 K, with a temperature difference of no more than 2 K between points.
- **Equilibrium Time:** No more than 15 seconds (this can be relaxed to 30 seconds if specific heating rate conditions are met).
- **Drying Performance:** Moisture content change of solid load ≤ 0.2%, porous load ≤ 1.0%.
- **Dynamic Pressure Change Rate:** No more than 10 bar/min to prevent packaging damage.
Instruments used in the test, such as **temperature sensors (platinum resistance thermometers conforming to EN IEC 60751)** and **pressure sensors**, must have high accuracy and fast response (τ0.9 < 1 second). **Recorders** can be analog or digital, independent of the control system.


Explanation of Technical Terms and Application Cases

Process Challenge Device (PCD):: A device used to simulate the most challenging sterilization conditions. For example, a narrow-cavity PCD contains a long, thin tube with a chemical or biological indicator inside. The indicator will only change color or become inactive when steam fully penetrates the tube. In dental clinics, when using a Type B sterilizer to process handpieces with lumens, validation using a narrow-cavity PCD is mandatory.
Theoretical Steam Temperature: The saturated steam temperature calculated by measuring pressure. Comparing this temperature with the measured temperature determines whether the steam quality is saturated. A large deviation indicates the presence of non-condensable gases or overheating, which will affect the sterilization effect.
Case Study: A beauty salon uses a Type N sterilizer only to process non-wrapped metal forceps, without requiring a drying stage; while a hospital's sterilization supply center uses a Type B sterilizer to process wrapped, multi-hole surgical drapes and lumens, requiring a full set of tests including air leak tests and multi-hole load drying tests.

Implementation Recommendations

For manufacturing companies, it is recommended to incorporate risk management and availability engineering (compliant with EN ISO 14971 and EN 62366-1) during the design phase to ensure software reliability and meet independent monitoring requirements. Users should provide compliant water, electricity, and steam (if externally supplied) according to the manufacturer's installation instructions and conduct regular leak tests and calibrations. Users are advised to select the appropriate sterilization cycle (B/N/S) based on the actual load type and conduct routine checks using preset test cycles (such as air leak test cycles). For locations handling lumen instruments, Type B sterilizers validated by PCD should be preferred.

3Version History

BS EN 13060:2004 older 2005-08-03
BS EN 13060:2005 older 2005-08-03
BS EN 13060+A1:2005 Amd A1/2005 older 2005-08-03
BS EN 13060+A2:2005 Amd A2/2005 older 2005-08-03
BS EN 13060:2004+A1:2009 Amd A1/2009 older 2005-08-03
BS EN 13060:2004+A2:2010 Amd A2/2010 older 2005-08-03
BS EN 13060:2014 older 2014-12-31
BS EN 13060:2014+A1:2018 Amd A1/2018 older 2014-01-01
BS EN 13060:2025 2025-08-31

5Citation Network

0
Cite this standard
21
Referenced herein
EN 10088-1:2023 EN 13445-3:2021 EN 13445-8:2021 EN 285:2015+A1:2021 EN 60529:1991 EN 61010-1:2010 EN 62366-1:2015 EN 764-7:2002

6Frequently Asked Questions

What is BS EN 13060:2025?
BS EN 13060:2025 — Sterilizers for medical purposes. Small steam sterilizers. Requirements and testing is an international standard developed by British Standards Institution (BSI). 1 Scope This document specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids....
What does BS EN 13060:2025 cover?
This standard covers: 1 Scope This document specifies the performance requirements and test methods for small steam sterilizers and sterilization cycles which are used for medical purposes or for materials that are likely to come into contact with blood or body fluids. This document applies to automatically controlled...
Who should use this standard?
This standard is intended for organizations, professionals, and stakeholders involved in various industries and sectors. It is applicable to manufacturers, service providers, regulatory bodies, and certification organizations.
What is the latest version of BS EN 13060:2025?
The current published version is BS EN 13060:2025, published on 2025-08-31. Always check for amendments or pending revisions.
How do I purchase BS EN 13060:2025?
You can purchase BS EN 13060:2025 by visiting our contact page. Click "Get This Standard" to view pricing and send a purchase inquiry. Our team will respond with payment and delivery details.